03.11.2013
On this video we are showing SiPAP using the old circuit. SiPAP Infant flow LP Circuit is now available to use,this is a new sipap circuit that gives more prominent pressure and spikes compare in using the old circuit. During our clinical demo we just found out the different pressure given by new circuit versus old circuit Starting the SiPAP 1. Attach the circuit properly make sure of no leak 2. Switch ON (BACK) 3.Set the low pressure to 9 cmH20 and high flow to 3cmH20 (initial setting only) 3.Do Calibration test (set FiO2 to 21% and 100%) checking the patency of oxygen cell then press exit 4.Do the Leak Test and alarm test (ncpap,fio2,highpressure,apnea transducer) 5.ncpap option will show only. 6. press ncpap ,other 4 option will be visible (ncpap apnea, biphasic, biphasic apnea, biphasic trigger) please see the video for more info..