PT vent assessment Part 2 HD

08.06.2012
SIMV VC, VT 400ml, RR 12, 21%, + 5 Peep, 2 PS Set I-Time (.9) Pause (0), Rise Time (.15), Flow Trigger set at 2, Insp Cycle Off (30%), Check Measured Values: PIP 33, Peep +5, RR (12) Fio2(21%), Ve 5.0L, VTi 399ml, VTe 413ml, Check Alarms: High Pressure (10cmH20 above, Ve Low 2l below set High Ve 2 or 3 above set Ve, Low Rate about 4 below set RR, High Rate about 5 above, Low PEEP alarm 2 below set PEEP, High PEEP alarm 2 above set PEEP